DEEPBLUE Fecal Occult Blood (FOB) Test Cassette (Colloidal Gold)


Batch (order inquiry: 📞84814841): instant query
Price:
Sale price$0.00

Description

Intended use

Fecal Occult Blood (FOB) Test (Colloidal Gold) is used for qualitative detection of hemoglobin in human feces at home for self-testing, or at all levels of medical institutions for testing, so as to provide auxiliary diagnosis for digestive tract haemorrhagic disease.



Test principle

This assay is a one-step lateral flow chromatographic immunoassay. The test strip consists of: 1) a burgundy colored conjugate pad containing mouse anti-hHb antibodies conjugated with collagenal gold and 2) antrocellular membrane strip containing a Test line (T- and a Control line (C-line). The T-line is coated with anti-hHb antibodies, and the C-line is coated with goat anti-mouse IgG antibodies. When an adequate volume of test specimen is dispensed into the sample well of the device, the test specimen migrates by capillary action across the test strip. If the concentration of hHb in the specimen is at or above 50 ng/ ml, the T-line appears as a visible burgundy line. If the concentration of hHb in the The specimen is below the detectable level, no T-line develops.




Materials and Components

1.20 test devices (cassettes), each sealed in a foil pouch.

2. Fecal sample tube including 2ml dilute buffer (0.85% NaCL) for each test.

3. One package insert (instructions for use).

4. Timer.

5. Fecal container.



Product Information


Specification 1pcs/box, 5pcs/box, 25pcs/box
Specimen Stool
Storage 4-30℃


Test Procedure

Bring all materials and specimens to room temperature (4-30°C)

1. Specimen Collection

Randomly pierce the fecal specimen with the swab or stick at least 5 different sites and put into the fecal collection tubes with buffer.( 1 . Fecal samples should be collected in clean and dry containers (no preservatives or transport media). 2 .Urines should be avoided to collect in the containers.


2. Testing

( 1 ) Shake the fecal collection tube vigorously to ensure a good liquid suspension.
( 2 ) Remove the test cassette from the sealed foil pouch and place it on a flat surface.

( 3 ) Dispense 3 or 4 drops processed samples into the sample well on the test (S).



3. Test Results


Read the result within 10 minutes. Do not interpret results after 10 minutes